Hioki pull-through AC/DC current sensors are best-in-class devices for use with power meters, Memory HiCorders and high performance oscilloscopes.
The CT6875 is a DC to 2MHz/500A sensor, ideal for capturing current signals from switching power supplies, inverters, power conditioners and other next generation devices. CAT III 1000 V
Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Rated current |
500 A AC/DC |
Max. allowable input |
Within the derating range, up to ±1500 Apeak (design value) allowed at 40°C or less for 20 ms or less |
Frequency bandwidth |
Amplitude: DC to 2 MHz (CT6875), DC to 1.5 MHz (CT6875-01)
Phase: DC to 1 MHz |
Basic accuracy |
(DC, 45 Hz ≤ f ≤ 66 Hz) Amplitude: ±0.04 % rdg. ±0.008 % f.s., Phase: ±0.1° |
Output voltage rate |
4 mV / A rated (This device outputs AC+DC voltage via the Sensor Unit. Use with voltage input equipment with high input impedance such as 1 MΩ) |
Max. rated voltage to earth |
1000 V AC/DC (50/60 Hz, CAT III ) |
Core diameter |
φ 36 mm (1.42 in) |
Operating temperature, humidity |
-40°C to +85°C (-40°F to 185°F), 80% RH or less (with no condensation) |
Power supply |
Power suppled via the Power Analyzer PW6001, PW3390, or Sensor Unit CT9555, CT9556, CT9557 |
Max. rated power |
7 VA |
Dimensions and mass |
160 mm (6.30 in)W × 112 mm (4.41 in)H × 50 mm (1.97 in)D, CT6875: 800 g (28.2 oz), cable length 3 m (9.84 ft), CT6875-01: 1100 g (38.8 oz), cable length 10 m (32.81 ft) |
Accessories |
Instruction manual ×1, Mark bands ×6, Operating precautions ×1 |
Compatible models
Compatible models |
CT6875 |
Power Analyzer PW6001, PW3390 |
Ok |
Power Analyzer 3390 |
Attention (Requires CT9901) |
Power HiTester 3193 series |
Attention (Requires CT9901) |
Current Unit 8971 |
Attention (Requires the 9318, CT9901) |